A nine-degree-of-freedom optimization-based method to evaluate the isocenter coincidence of the treatment beams and image system of a medical accelerator

التفاصيل البيبلوغرافية
العنوان: A nine-degree-of-freedom optimization-based method to evaluate the isocenter coincidence of the treatment beams and image system of a medical accelerator
المؤلفون: Yun Zhou, Jian Qiao, Nan Yan, Liyan Dai, Yuehu Pu
المصدر: Frontiers in Physics, Vol 12 (2024)
بيانات النشر: Frontiers Media S.A., 2024.
سنة النشر: 2024
المجموعة: LCC:Physics
مصطلحات موضوعية: radiation therapy (radiotherapy), medical accelerator, isocenter coincidence, quality assurance, cone-beam computed tomography (CBCT), Physics, QC1-999
الوصف: Medical accelerators have been widely used in tumor radiation therapy. Accurate isocenter coincidence between treatment beams and imaging systems is critical for image-guided radiation therapy (IGRT). We propose a method utilizing a phantom with marker spheres to detect the Nine Degrees of Freedom (9-DOF) in the system’s geometric model to assess isocenter coincidence between the treatment beams and the kV cone-beam computed tomography (CBCT). The phantom was initially aligned with the accelerator. Subsequently, the projections of the treatment and CBCT beams’ were acquired separately with full gantry rotation. By analyzing the marker spheres’ position in both the treatment beam and CBCT beam projections, the 9-DOF parameters were calculated. A comparison with a Winston-Lutz-based system was performed. Then, the analysis revealed imprecise circular trajectories with noticeable random deviations in the rotations of both the treatment beams and CBCT. The isocenter deviations for the treatment beams and CBCT were 0.18 mm (X), −0.49 mm (Y), and −0.35 mm (Z) after trajectories fitting, respectively. The rotational planes of the two systems exhibited a pinch angle of 0.0235°. This proposed method offers a quantitative assessment of the geometric pose of the source and the detector panel, and the isocenter coincidence of the treatment beams and imaging systems of an accelerator at each gantry angle.
نوع الوثيقة: article
وصف الملف: electronic resource
اللغة: English
تدمد: 2296-424X
العلاقة: https://www.frontiersin.org/articles/10.3389/fphy.2024.1395997/fullTest; https://doaj.org/toc/2296-424XTest
DOI: 10.3389/fphy.2024.1395997
الوصول الحر: https://doaj.org/article/e4d7a4a6c8a545439975a4726c3ef869Test
رقم الانضمام: edsdoj.4d7a4a6c8a545439975a4726c3ef869
قاعدة البيانات: Directory of Open Access Journals
الوصف
تدمد:2296424X
DOI:10.3389/fphy.2024.1395997