Implementing and evaluating a high‐resolution diode array for patient‐specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy

التفاصيل البيبلوغرافية
العنوان: Implementing and evaluating a high‐resolution diode array for patient‐specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy
المؤلفون: Xu, Qianyi, Huynh, Kiet, Nie, Wei, Rose, Mark S., Chawla, Ashish K., Choe, Kevin S., Kanani, Samir, Kubicek, Gregory J., Fan, Jiajin
المصدر: Journal of Applied Clinical Medical Physics ; volume 23, issue 5 ; ISSN 1526-9914 1526-9914
بيانات النشر: Wiley
سنة النشر: 2022
المجموعة: Wiley Online Library (Open Access Articles via Crossref)
الوصف: The purpose of the study was to introduce and evaluate a high‐resolution diode array for patient‐specific quality assurance (PSQA) of CyberKnife brain stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). Thirty‐three intracranial plans were retrospectively delivered on the SRS MapCHECK using fixed cone, Iris, and multileaf collimator (MLC). The plans were selected to cover a range of sites from large tumor bed, single/multiple small brain metastases (METs) to trigeminal neuralgia. Fiducial tracking using the four fiducials embedded around the detector plane was used as image guidance. Results were analyzed before and after registration based on absolute dose gamma criterion of 1 mm distance‐to‐agreement and 0.5%–3% dose‐difference. Overall, the gamma passing rates (1 mm and 3% criterion) before registration for all the patients were above 90% for all three treatment modalities (96.8 ± 3.5%, the lowest passing rate of 90.4%), and were improved after registration (99.3 ± 1.5%). When tighter criteria (1 mm and 2%) were applied, the gamma passing rates after registration for all the cases dropped to 97.3 ± 3.2%. For trigeminal neuralgia cases, we applied 1 mm and 0.5% criterion and the passing rates dropped from 100 ± 0.0% to 98.5 ± 2.0%. The mean delivery time was 33.4 ± 11.7 min, 24.0 ± 4.9 min, and 17.1 ± 2.6 min for the fixed cone, Iris, and MLC, respectively. With superior gamma passing rates and reasonable quality assurance (QA) time, we believe the SRS MapCHECK could be a good option for routine PSQA for CyberKnife SRS/SRT.
نوع الوثيقة: article in journal/newspaper
اللغة: English
DOI: 10.1002/acm2.13569
الإتاحة: https://doi.org/10.1002/acm2.13569Test
حقوق: http://creativecommons.org/licenses/by/4.0Test/
رقم الانضمام: edsbas.D7AA9088
قاعدة البيانات: BASE