العنوان: Investigation of multifilament MgB2 superconducting joint technique for development of MRI magnets
المؤلفون: Yoo, B. H., Kim, J. C., Kim, Y. G., Hwang, D. Y., Lee, J. H., Lee, H. G.
المساهمون: Korea Evaluation Institute of Industrial Technology
المصدر: Review of Scientific Instruments ; volume 89, issue 9 ; ISSN 0034-6748 1089-7623
بيانات النشر: AIP Publishing
سنة النشر: 2018
مصطلحات موضوعية: Instrumentation
الوصف: This study presents the investigation of superconducting joints fabricated using multifilament magnesium diboride (MgB2) wires for the development of persistent-current mode magnetic resonance imaging (MRI) magnets. The critical current of the jointed samples decreased with increasing cutting angle because the smaller cutting angle allowed greater exposure of the MgB2 filament, thereby increasing the contact area for the wire-bulk-wire connection. In addition, an appropriate pressing pressure (300 MPa) was necessary to establish the multifilament MgB2 joint without significant degradation of superconducting properties. The resistance of the optimal MgB2 joint, measured using the field-decay technique, was <1.5 × 10−14 Ω. Therefore, the proposed joint technique can be employed for developing multifilament MgB2 MRI magnets operating in the persistent-current mode.
نوع الوثيقة: article in journal/newspaper
اللغة: English
DOI: 10.1063/1.5040549
DOI: 10.1063/1.5040549/16140793/094701_1_online.pdf
الإتاحة: https://doi.org/10.1063/1.5040549Test