Measurement of scintillation response of CsI[Na] to low-energy nuclear recoils by COHERENT

التفاصيل البيبلوغرافية
العنوان: Measurement of scintillation response of CsI[Na] to low-energy nuclear recoils by COHERENT
المؤلفون: Akimov, D., An, P., Awe, C., Barbeau, P. S., Becker, B., Belov, V., Bernardi, I., Blackston, M. A., Bock, C., Bolozdynya, A., Browning, J., Cabrera-Palmer, B., Chernyak, D., Conley, E., Daughhetee, J., Detwiler, J., Ding, K., Durand, M. R., Efremenko, Y., Elliott, S. R., Fabris, L., Febbraro, M., Rosso, A. Gallo, Galindo-Uribarri, A., Green, M. P., Heath, M. R., Hedges, S., Hoang, D., Hughes, M., Johnson, T., Khromov, A., Konovalov, A., Kozlova, E., Kumpan, A., Li, L., Link, J. M., Liu, J., Mann, K., Markoff, D. M., Mastroberti, J., Melikyan, Y. A., Mueller, P. E., Newby, J., Parno, D. S., Penttila, S. I., Pershey, D., Rapp, R., Ray, H., Raybern, J., Razuvaeva, O., Reyna, D., Rich, G. C., Ross, J., Rudik, D., Runge, J., Salvat, D. J., Salyapongse, A. M., Scholberg, K., Shakirov, A., Simakov, G., Sinev, G., Snow, W. M., Sosnovtsev, V., Suh, B., Tayloe, R., Tellez-Giron-Flores, K., Tolstukhin, I., Ujah, E., Vanderwerp, J., Varner, R. L., Virtue, C. J., Visser, G., Wongjirad, T., Yen, Y. -r., Yoo, Jonghee, Yu, C. -h., Zettlemoyer, J., COHERENT Collaboration, C. O. H. E. R. E. N. T. Collaboration
المساهمون: Yoo, Jonghee
بيانات النشر: Institute of Physics
سنة النشر: 2022
المجموعة: Seoul National University: S-Space
الوصف: We present results of several measurements of CsI[Na] scintillation response to 3-60 keV energy nuclear recoils performed by the COHERENT collaboration using tagged neutron elastic scattering experiments and an endpoint technique. Earlier results, used to estimate the coherent elastic neutrino-nucleus scattering (CEvNS) event rate for the first observation of this process achieved by COHERENT at the Spallation Neutron Source (SNS), have been reassessed. We discuss corrections for the identified systematic effects and update the respective uncertainty values. The impact of updated results on future precision tests of CEvNS is estimated. We scrutinize potential systematic effects that could affect each measurement. In particular we confirm the response of the H11934-200 Hamamatsu photomultiplier tube (PMT) used for the measurements presented in this study to be linear in the relevant signal scale region. ; Y ; 1
نوع الوثيقة: article in journal/newspaper
اللغة: unknown
تدمد: 1748-0221
العلاقة: Journal of Instrumentation, Vol.17 No.10, p. P10034; https://hdl.handle.net/10371/188888Test; 000882475300007; 2-s2.0-85141225733; 176440
DOI: 10.1088/1748-0221/17/10/P10034
الإتاحة: https://doi.org/10.1088/1748-0221/17/10/P10034Test
https://hdl.handle.net/10371/188888Test
رقم الانضمام: edsbas.745CB859
قاعدة البيانات: BASE
الوصف
تدمد:17480221
DOI:10.1088/1748-0221/17/10/P10034