Effect of audiological testing on programmable ventriculoperitoneal shunts
العنوان: | Effect of audiological testing on programmable ventriculoperitoneal shunts |
---|---|
المؤلفون: | Ayeshah Abdul-Hamid, Gulam Zilani, Catriona Bryant, Samuel MacKeith, Alan Bryant, Ali Qureishi, Sanjeeva Jeyaretna |
المصدر: | International Journal of Audiology. 61:173-176 |
بيانات النشر: | Informa UK Limited, 2021. |
سنة النشر: | 2021 |
مصطلحات موضوعية: | Linguistics and Language, medicine.medical_specialty, business.industry, Transducers, Testing equipment, Equipment Design, Audiology, Ventriculoperitoneal Shunt, Language and Linguistics, 03 medical and health sciences, Speech and Hearing, Ventriculoperitoneal shunts, 0302 clinical medicine, Acoustic Impedance Tests, Humans, Medicine, 030223 otorhinolaryngology, business, Audiology equipment, 030217 neurology & neurosurgery |
الوصف: | This study aimed at measuring the magnetic field strength of commonly used types of audiological testing equipment and determine their effects on the three most commonly used programmable ventriculoperitoneal (VP) shunts to try and quantify the risk of a VP shunt being reprogrammed during audiological testing.In thisMagnetic field strength associated with transducers placed on a model of a skulls having implanted Miethke ProGAV 2.0, Medtronic Strata II and Codman Hakim programmable VP shunts was measured.The supra-aural earphones had a magnetic field strength of 14 mT at 0 mm, which dropped to 0 mT at 10 mm away from the transducer. All other equipment had a magnetic field strength of 3.5 mT or less at 0 mm. There was no instance of reprogramming of the shunts by the transducers.The findings suggest that the risk of inadvertent valve-reprogramming by the transducers is extremely small. However, care should be taken to avoid placing any of the transducers directly over the shunt. |
تدمد: | 1708-8186 1499-2027 |
الوصول الحر: | https://explore.openaire.eu/search/publication?articleId=doi_dedup___::f3482212121993c94200396310db7dafTest https://doi.org/10.1080/14992027.2021.1881174Test |
رقم الانضمام: | edsair.doi.dedup.....f3482212121993c94200396310db7daf |
قاعدة البيانات: | OpenAIRE |
تدمد: | 17088186 14992027 |
---|