Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor

التفاصيل البيبلوغرافية
العنوان: Accuracy of a Fourth-Generation Subcutaneous Continuous Glucose Sensor
المؤلفون: Bruce W. Bode, Mark P. Christiansen, Robert H. Slover, Ashley Sullivan, Suiying Huang, Ronald L. Brazg, Francine R. Kaufman, Scott W. Lee, John H. Shin, Timothy S. Bailey, Satish K. Garg
المصدر: Diabetes technologytherapeutics. 19(8)
سنة النشر: 2017
مصطلحات موضوعية: Insulin pump, Adult, Blood Glucose, Male, medicine.medical_specialty, Adolescent, Endocrinology, Diabetes and Metabolism, 030209 endocrinology & metabolism, Type 2 diabetes, 03 medical and health sciences, Young Adult, 0302 clinical medicine, Endocrinology, Subcutaneous glucose sensor, Insulin Infusion Systems, Reference Values, Diabetes mellitus, Commentaries, medicine, Fourth generation, Humans, 030212 general & internal medicine, Aged, business.industry, Blood Glucose Self-Monitoring, Reproducibility of Results, Middle Aged, medicine.disease, Surgery, Medical Laboratory Technology, Diabetes Mellitus, Type 2, Reference values, Anesthesia, Calibration, Female, business
الوصف: This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (GuardianEighty-eight subjects (14-75 years of age, mean ± standard deviation [SD] of 42.0 ± 19.1 years) with type 1 or type 2 diabetes participated in the study. Subjects wore two sensors in the abdomen that were paired with either a MiniMedThe overall mean absolute relative difference (MARD ± SD) between abdomen sensor glucose (SG) and YSI reference values was 9.6% ± 9.0% and 9.4% ± 9.8% for the MiniMed 640G insulin pump and Guardian Connect system, respectively; and 8.7% ± 8.0% between arm SG and YSI reference values. The percentage of SG values within 20% agreement of the YSI reference value (for YSI80 mg/dL) was 90.7% with the MiniMed 640G insulin pump, 91.8% with the Guardian Connect system, and 93.1% for GSR-connected arm sensors. Mean functional sensor life, when calibrating 3-4 times/day, was 145.9 ± 39.3 h for sensors paired with the MiniMed 640G insulin pump, 146.1 ± 41.6 h for sensors paired with the Guardian Connect system, and 147.6 ± 40.4 h for sensors connected to the GSR. Responses to survey questions regarding sensor comfort and ease of use were favorable.The Guardian Sensor 3 glucose sensor, whether located in abdomen or the arm, provided accurate glucose readings when compared with the YSI reference and demonstrated functional life commensurate with the intended 7-day use. ClinicalTrials.gov : NCT02246582.
تدمد: 1557-8593
الوصول الحر: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::ab962a052b91ea68df807aa7509e5942Test
https://pubmed.ncbi.nlm.nih.gov/28767273Test
حقوق: OPEN
رقم الانضمام: edsair.doi.dedup.....ab962a052b91ea68df807aa7509e5942
قاعدة البيانات: OpenAIRE