Effect of exposure from iPhone 12 on programmable ventriculoperitoneal shunts

التفاصيل البيبلوغرافية
العنوان: Effect of exposure from iPhone 12 on programmable ventriculoperitoneal shunts
المؤلفون: Ameeta Kumar, Arbaaz Pervaiz, Anouk Borg, Ayeshah Abdul-Hamid, Sanjeeva Jeyaretna, Samuel MacKeith, Ali Qureishi
المصدر: British Journal of Neurosurgery. 36:415-419
بيانات النشر: Informa UK Limited, 2022.
سنة النشر: 2022
مصطلحات موضوعية: Humans, Surgery, Equipment Design, Prostheses and Implants, Neurology (clinical), General Medicine, Ventriculoperitoneal Shunt, Cerebrospinal Fluid Shunts, Hydrocephalus
الوصف: The latest iPhone 12 model has elicited concerns over its interaction with medical devices such as pacemakers due to its integrated MagSafe technology. Historically, programmable ventriculoperitoneal (VP) shunts have been demonstrated to readjust when exposed to magnetic objects. Yet, the presence of interactions between the iPhone 12 and shunts is unknown. In this in-vitro study, we examined the effect on the programming of three VP shunts, Medtronic Strata II, Miethke ProGAV 2.0 and Codman Hakim, when exposed to the iPhone 12 model. We found that all three valves did not re-program when the iPhone was held near or moved in a swiping or rotational motion above the valves. Therefore, the risk of re-programming of these three shunts when exposed to iPhone 12 appears to be low. However, patients should take care until further work is undertaken to examine the complex interplay between programmable VP shunts with magnetic devices.
تدمد: 1360-046X
0268-8697
الوصول الحر: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::913cdab72868cab9080491761e17f2a9Test
https://doi.org/10.1080/02688697.2022.2028724Test