Remote Continuous Glucose Monitoring With a Computerized Insulin Infusion Protocol for Critically Ill Patients in a COVID-19 Medical ICU: Proof of Concept

التفاصيل البيبلوغرافية
العنوان: Remote Continuous Glucose Monitoring With a Computerized Insulin Infusion Protocol for Critically Ill Patients in a COVID-19 Medical ICU: Proof of Concept
المؤلفون: Guillermo E. Umpierrez, Georgia Davis, Barbara McLean, Shivani Agarwal, Kathleen Dungan, Citlalli Perez-Guzman, Francisco J. Pasquel, Marina Rabinovich, Limin Peng, Eileen R Faulds, Norma Poindexter, Tara Walker, Joi C. Hester, Debbie Vigliotti, Patricia Hannon, Seema S. Tekwani, Petrena Saunders, Greg S. Martin
المصدر: Diabetes Care
بيانات النشر: American Diabetes Association, 2021.
سنة النشر: 2021
مصطلحات موضوعية: Blood Glucose, Male, medicine.medical_specialty, Coronavirus disease 2019 (COVID-19), Critical Illness, Point-of-Care Systems, Endocrinology, Diabetes and Metabolism, medicine.medical_treatment, 030209 endocrinology & metabolism, Proof of Concept Study, law.invention, Diabetes Complications, 03 medical and health sciences, Insulin Infusion Systems, 0302 clinical medicine, law, Diabetes mellitus, Internal Medicine, medicine, Humans, Insulin, 030212 general & internal medicine, Aged, Aged, 80 and over, Advanced and Specialized Nursing, Mechanical ventilation, Protocol (science), SARS-CoV-2, Continuous glucose monitoring, business.industry, Critically ill, Blood Glucose Self-Monitoring, COVID-19, Middle Aged, medicine.disease, Intensive care unit, COVID-19 Drug Treatment, Novel Communications in Diabetes, Clinical trial, Intensive Care Units, Equipment and Supplies, Remote Sensing Technology, Emergency medicine, Female, business, Algorithms
الوصف: OBJECTIVE The use of remote real-time continuous glucose monitoring (CGM) in the hospital has rapidly emerged to preserve personal protective equipment and reduce potential exposures during coronavirus disease 2019 (COVID-19). RESEARCH DESIGN AND METHODS We linked a hybrid CGM and point-of-care (POC) glucose testing protocol to a computerized decision support system for continuous insulin infusion and integrated a validation system for sensor glucose values into the electronic health record. We report our proof-of-concept experience in a COVID-19 intensive care unit. RESULTS All nine patients required mechanical ventilation and corticosteroids. During the protocol, 75.7% of sensor values were within 20% of the reference POC glucose with an associated average reduction in POC of 63%. Mean time in range (70–180 mg/dL) was 71.4 ± 13.9%. Sensor accuracy was impacted by mechanical interferences in four patients. CONCLUSIONS A hybrid protocol integrating real-time CGM and POC is helpful for managing critically ill patients with COVID-19 requiring insulin infusion.