Myopotential Inhibition of Unipolar Lithium Pacemakerss

التفاصيل البيبلوغرافية
العنوان: Myopotential Inhibition of Unipolar Lithium Pacemakerss
المؤلفون: Levine, Paul A., Caplan, Charles H., Klein, Michael D., Brodsky, Samuel J., Ryan, Thomas J.
المصدر: Chest; October 1982, Vol. 82 Issue: 4 p461-465, 5p
مستخلص: The effect of Isometric upper extremity exercise on pacemaker function was evaluated in 27 patients who remained pacemaker-dependent during testing. Inhibition was demonstrated in eight (31 percent) of which five were symptomatic. Based on design of the sensing amplifier and return to an all-metal housing in the unipolar lithium pulse generators, myopotential inhibition is being recognized as one cause of symptomatic pacemaker inhibition that is more common than generally appreciated. A method of evaluation and management options for symptomatic patients are discussed. Routine testing of all patients should be performed at the time of a regular office evaluation. If one model pulse generator appears to be particularly prone to myopotential inhibition, this knowledge should be considered in the choice of future pacing systems.
قاعدة البيانات: Supplemental Index
الوصف
تدمد:00123692
19313543
DOI:10.1378/chest.82.4.461