An in vitroassessment of wound dressing conformability

التفاصيل البيبلوغرافية
العنوان: An in vitroassessment of wound dressing conformability
المؤلفون: Queen, D, Evans, J.H, Gaylor, J.D.S, Courtney, J.M, Reid, W.H
المصدر: Biomaterials; September 1987, Vol. 8 Issue: 5 p372-376, 5p
مستخلص: An in vitroassessment technique has been developed to determine the conformability of wound dressings. The technique employed is based on an inflation technique which provides a measurement of the minimum radius of curvature which a specific dressing will adopt under pressure. A pressure of 40 mmHg was chosen as this had been shown to be the maximum tolerable pressure before the occurrence of tissue breakdown. This radius is then matched to the natural radii of the body surfaces and an assessment of conformability can be made.
قاعدة البيانات: Supplemental Index
الوصف
تدمد:01429612
DOI:10.1016/0142-9612(87)90008-1