التفاصيل البيبلوغرافية
العنوان: |
Assessment of the dosimetric accuracies of CATPhan 504 and CIRS 062 using kV-CBCT for performing direct calculations. |
المؤلفون: |
Annkah, James Kwame1,2 jannkah@hotmail.com, Rosenberg, Ivan1,3, Hindocha, Naina3, Moinuddin, Syed Ali1, Ricketts, Kate1, Adeyemi, Abiodun4, Royle, Gary1 |
المصدر: |
Journal of Medical Physics. Jul-Sep2014, Vol. 39 Issue 3, p133-141. 9p. |
مصطلحات موضوعية: |
*RADIOTHERAPY, *CONE beam computed tomography, *RADIATION dosimetry, *RADIATION measurements, *MEDICAL physics |
مستخلص: |
The dosimetric accuracies of CATPhan 504 and CIRS 062 have been evaluated using the kV-CBCT of Varian TrueBeam linac and Eclipse TPS. The assessment was done using the kV-CBCT as a standalone tool for dosimetric calculations towards Adaptive replanning. Dosimetric calculations were made without altering the HU-ED curves of the planning computed tomography (CT) scanner that is used by the Eclipse TPS. All computations were done using the images and dataset from kV-CBCT while maintaining the HU-ED calibration curve of the planning CT (pCT), assuming pCT was used for the initial treatment plan. Results showed that the CIRS phantom produces doses within ±5% of the CT-based plan while CATPhan 504 produces a variation of ±14% of the CT-based plan. [ABSTRACT FROM AUTHOR] |
قاعدة البيانات: |
Academic Search Index |