A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays.

التفاصيل البيبلوغرافية
العنوان: A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays.
المؤلفون: Boissier, Elodie, Sévin-Allouet, Mathieu, Le Thuaut, Aurélie, De Gaalon, Solène, Trossaërt, Marc, Rozec, Bertrand, Lakhal, Karim, Rigal, Jean-Christophe
المصدر: Clinical Chemistry & Laboratory Medicine; Jun2017, Vol. 55 Issue 6, pe18-e21, 4p, 1 Chart, 1 Graph
مصطلحات موضوعية: HEMOSTASIS, CENTRIFUGATION, BLOOD coagulation tests, BLOOD testing, IMMUNOLOGICAL blood tests
مستخلص: The article presents a study which proves that a 2-minute at 4500 relative centrifugal force (g) rather than a 15-minutes at 2200 g centrifugation does not impact the reliability of coagulation assays. It mentions that the findings have a positive impact on patients' outcome in emergent situations where a protracted turnaround time (TAT) of hemostasis tests may delay urgent therapies.
قاعدة البيانات: Complementary Index
الوصف
تدمد:14346621
DOI:10.1515/cclm-2016-0670