A specific absorption rate reduction method for simultaneous multislice magnetic resonance imaging.

التفاصيل البيبلوغرافية
العنوان: A specific absorption rate reduction method for simultaneous multislice magnetic resonance imaging.
المؤلفون: Fu-Hsing Wu, Wu, Edzer L., Yi-Hang Tung, Po-Wei Cheng, Tzi-Dar Chiueh, Jyh-Horng Chen
المصدر: Review of Scientific Instruments; Apr2017, Vol. 88 Issue 4, p1-12, 12p, 3 Diagrams, 1 Chart, 5 Graphs
مصطلحات موضوعية: ABSORPTION, MAGNETIC resonance imaging, RADIO frequency, ALGORITHMS, BRAIN imaging, IMAGE quality analysis
مستخلص: This study proposes a modified Shinnar-Le Roux method to synthesize the excitation radio frequency (RF) pulse for a 2D gradient echo (GRE) based simultaneous multi-slice (SMS) magnetic resonance imaging (MRI) with features of low specific absorption rate (SAR) and small out-of-slice ripple. This synthesis method for SMS RF pulses employs thinner slice bandwidth and lower multislice offset frequencies to reduce SAR values and adopts a weighted Parks-McClellan algorithm to reduce sidelobes. Formulas for estimating relative SAR values of the SMS pulses are also introduced. Relative SAR values and out-of-slice ripples of the proposed and typical RF pulses with different parameters are presented. In simultaneous 5-slice phantom and 3-slice human brain imaging, SMS pulses synthesized with the proposed method achieve 32% and 28% SAR values of standard pulses while providing similar image qualities. Typical RF pulses such as sinc x cos can also take advantage of the proposed method and offer lower SAR values for SMS imaging. The RF pulse synthesized using the proposed method features low SAR, small sidelobes, and consistent image quality for 2D GRE-based SMS MRI. This method is applicable to the synthesis of typical SMS RF pulses for significant SAR reduction. [ABSTRACT FROM AUTHOR]
Copyright of Review of Scientific Instruments is the property of American Institute of Physics and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:00346748
DOI:10.1063/1.4979861