Excimer laser (308 nm) recanalisation of in-stent restenosis: thermal considerations.

التفاصيل البيبلوغرافية
العنوان: Excimer laser (308 nm) recanalisation of in-stent restenosis: thermal considerations.
المؤلفون: Papaioannou, T., Yadegar, D., Vari, S., Shehada, R., Grundfest, W.S.
المصدر: Lasers in Medical Science; Mar2001, Vol. 16 Issue 2, p90-100, 11p
مستخلص: Excimer laser recanalisation of in-stent restenosis may be a viable modality for improving coronary patency. However, the presence of arterial stents modifies the thermal properties of the irradiated area and may alter temperature patterns generated during ablation. The goal of this study was to evaluate, in vitro, temperature changes during excimer laser ablation of stented vessels and compare them with those obtained from unstented (control) vessels. Six different stent types (AVE Microstent-II, AVE-GFX, ACS Multi-link, JJ Palmaz-Schatz, JJ Crown, and NIR) were deployed in freshly excised porcine coronary vessels. Three control unstented samples were also measured. Blood or saline was infused through the vessels, while the tissue environment was kept at approximately 37 degrees C. A 308 nm excimer laser (Spectranetics, CVX300) with an eccentric 2.0 mm laser catheter (Spectranetics, EII) delivered two trains of 200 pulses each, 10 s apart, at 60 mJ/mm2, and 40 Hz, simulating maximum clinical exposure. The catheter was positioned midway in the stent, first coaxially parallel to the vessel wall, and then at an angle against the stent and vessel wall. Temperature measurements (n= 168 for blood, n=96 for saline) were performed with a approximately 210 microm diameter, fast-response thermocouple with 0.1 degrees C resolution. The probe was positioned to within approximately 250 microm from the inner surface of the vessels. Tissue temperature was measured at the catheter tip and at the distal and proximal edges of the stents. Maximum recorded temperatures for coaxial and angular alignment, did not exceed 42.2 degrees C (approximately 6 degrees C above baseline) and 54.2 degrees C (approximately 18.1 degrees C above baseline) respectively, for all stents types tested, controls, and all probe locations. Both stented and unstented vessels exhibited comparable temperature gradients. The observed maximum temperatures, obtained under extreme lasing conditions, indicated that 308 nm ablation, in the presence of stents under blood or saline infusion, produces clinically acceptable temperatures. [ABSTRACT FROM AUTHOR]
Copyright of Lasers in Medical Science is the property of Springer Nature and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:02688921
DOI:10.1007/PL00011348