Comparison of Microlife BP A200 Plus and Omron M6 blood pressure monitors to detect atrial fibrillation in hypertensive patients

التفاصيل البيبلوغرافية
العنوان: Comparison of Microlife BP A200 Plus and Omron M6 blood pressure monitors to detect atrial fibrillation in hypertensive patients
المؤلفون: Francesco Pelliccia, Maurizio Volterrani, Giuseppe Marazzi, Fabrizia Grieco, Luca Cacciotti, Giuseppe Caminiti, Mauro Lombardo, F. Iellamo, Giuseppe M.C. Rosano, Daniela Righi, Luigi Iaia
المصدر: Advances in therapy. 29(1)
سنة النشر: 2011
مصطلحات موضوعية: Male, medicine.medical_specialty, Settore MED/09 - Medicina Interna, Blood Pressure, Essential hypertension, self-blood pressure measurement, Electrocardiography, Blood Pressure Monitoring, Heart Rate, Predictive Value of Tests, Internal medicine, Heart rate, Ambulatory, Atrial Fibrillation, medicine, Humans, Pharmacology (medical), Prospective Studies, Prospective cohort study, Aged, Microlife, medicine.diagnostic_test, business.industry, Medicine (all), Atrial fibrillation, General Medicine, Blood Pressure Monitoring, Ambulatory, Middle Aged, medicine.disease, Blood pressure, Early Diagnosis, atrial fibrillation, diagnostic accuracy, hypertension, Female, Hypertension, Predictive value of tests, Cardiology, business
الوصف: Self-monitoring home blood pressure (BP) devices are currently recommended for long-term follow-up of hypertension and its management. Some of these devices are integrated with algorithms aimed at detecting atrial fibrillation (AF), which is common essential hypertension. This study was designed to compare the diagnostic accuracy of two widely diffused home BP monitoring devices in detecting AF in an unselected population of outpatients referred to a hypertension clinic because of high BP.In 503 consecutive patients the authors simultaneously compared the accuracy of the Microlife(®) BP A200 Plus (Microlife) and the OMRON(®) M6 (OMRON) home BP devices, in detecting AF.Systolic and diastolic BP as well as heart rate (HR) values detected by the two devices were not significantly different. Pulse irregularity was detected in 124 and 112 patients with the OMRON M6 and Microlife BP A200 Plus devices, respectively. Simultaneous electrocardiogram (ECG) recording revealed that pulse irregularity was due to AF in 101 patients. Pulse irregularity detected by the OMRON M6 device corresponded to AF in 101, to supraventricular premature beats in 18, and to frequent premature ventricular beat in five patients, respectively. Pulse irregularity detected by the Microlife BP A200 Plus device corresponded to AF in 93, to supraventricular premature beats in 14, and to ventricular premature beats in five patients. The sensitivity for detecting AF was 100%, the specificity was 92%, and diagnostic accuracy 95% for the OMRON M6 and 100%, 92%, and 95 for the Microlife BP A200 Plus, respectively. AF was newly diagnosed by ECG recordings in 47 patients, and was detected in all patients by the OMRON device, and in 42 patients by the Microlife device.These results indicate that OMRON M6 is more accurate than Microlife BP A200 Plus in detecting AF in patients with essential hypertension. Widespread use of these devices in hypertensive patients could be of clinical benefit for the early diagnosis and treatment of this arrhythmia and related consequences.
تدمد: 1865-8652
الوصول الحر: https://explore.openaire.eu/search/publication?articleId=doi_dedup___::3bb421a0688424d087dbce0f9a4eb64eTest
https://pubmed.ncbi.nlm.nih.gov/22198902Test
حقوق: OPEN
رقم الانضمام: edsair.doi.dedup.....3bb421a0688424d087dbce0f9a4eb64e
قاعدة البيانات: OpenAIRE