A 31T split-pair pulsed magnet for single crystal x-ray diffraction at low temperature

التفاصيل البيبلوغرافية
العنوان: A 31T split-pair pulsed magnet for single crystal x-ray diffraction at low temperature
المؤلفون: Duc, F., Fabrèges, X., Roth, T., Detlefs, C., Frings, P., Nardone, M., Billette, J., Lesourd, M., Zhang, L., Zitouni, A., Delescluse, P., Béard, J., Nicolin, J. P., Rikken, G. L. J. A.
المصدر: Rev. Sci. Instrum. 85, 053905 (2014)
سنة النشر: 2014
المجموعة: Condensed Matter
Physics (Other)
مصطلحات موضوعية: Condensed Matter - Strongly Correlated Electrons, Physics - Instrumentation and Detectors
الوصف: We have developed a pulsed magnet system with panoramic access for synchrotron x-ray diffraction in magnetic fields up to 31T and at low temperature down to 1.5 K. The apparatus consists of a split-pair magnet, a liquid nitrogen bath to cool the pulsed coil, and a helium cryostat allowing sample temperatures from 1.5 up to 250 K. Using a 1.15MJ mobile generator, magnetic field pulses of 60 ms length were generated in the magnet, with a rise time of 16.5 ms and a repetition rate of 2 pulses/hour at 31 T. The setup was validated for single crystal diffraction on the ESRF beamline ID06.
نوع الوثيقة: Working Paper
DOI: 10.1063/1.4878915
الوصول الحر: http://arxiv.org/abs/1403.0722Test
رقم الانضمام: edsarx.1403.0722
قاعدة البيانات: arXiv