دورية أكاديمية

T1 mapping performance and measurement repeatability: results from the multi-national T1 mapping standardization phantom program (T1MES).

التفاصيل البيبلوغرافية
العنوان: T1 mapping performance and measurement repeatability: results from the multi-national T1 mapping standardization phantom program (T1MES).
المؤلفون: Captur, Gabriella, Bhandari, Abhiyan, Brühl, Rüdiger, Ittermann, Bernd, Keenan, Kathryn E., Yang, Ye, Eames, Richard J., Benedetti, Giulia, Torlasco, Camilla, Ricketts, Lewis, Boubertakh, Redha, Fatih, Nasri, Greenwood, John P., Paulis, Leonie E. M., Lawton, Chris B., Bucciarelli-Ducci, Chiara, Lamb, Hildo J., Steeds, Richard, Leung, Steve W., Berry, Colin
المصدر: Journal of Cardiovascular Magnetic Resonance (BioMed Central); 5/7/2020, Vol. 22 Issue 1, p1-17, 17p, 3 Color Photographs, 1 Diagram, 4 Charts, 1 Graph
مصطلحات موضوعية: COMPARATIVE studies, MAGNETIC resonance imaging, MYOCARDIUM, IMAGING phantoms, QUALITY assurance, TEMPERATURE
مستخلص: Background: The T1 Mapping and Extracellular volume (ECV) Standardization (T1MES) program explored T1 mapping quality assurance using a purpose-developed phantom with Food and Drug Administration (FDA) and Conformité Européenne (CE) regulatory clearance. We report T1 measurement repeatability across centers describing sequence, magnet, and vendor performance. Methods: Phantoms batch-manufactured in August 2015 underwent 2 years of structural imaging, B0 and B1, and "reference" slow T1 testing. Temperature dependency was evaluated by the United States National Institute of Standards and Technology and by the German Physikalisch-Technische Bundesanstalt. Center-specific T1 mapping repeatability (maximum one scan per week to minimum one per quarter year) was assessed over mean 358 (maximum 1161) days on 34 1.5 T and 22 3 T magnets using multiple T1 mapping sequences. Image and temperature data were analyzed semi-automatically. Repeatability of serial T1 was evaluated in terms of coefficient of variation (CoV), and linear mixed models were constructed to study the interplay of some of the known sources of T1 variation. Results: Over 2 years, phantom gel integrity remained intact (no rips/tears), B0 and B1 homogenous, and "reference" T1 stable compared to baseline (% change at 1.5 T, 1.95 ± 1.39%; 3 T, 2.22 ± 1.44%). Per degrees Celsius, 1.5 T, T1 (MOLLI 5s(3s)3s) increased by 11.4 ms in long native blood tubes and decreased by 1.2 ms in short post-contrast myocardium tubes. Agreement of estimated T1 times with "reference" T1 was similar across Siemens and Philips CMR systems at both field strengths (adjusted R2 ranges for both field strengths, 0.99–1.00). Over 1 year, many 1.5 T and 3 T sequences/magnets were repeatable with mean CoVs < 1 and 2% respectively. Repeatability was narrower for 1.5 T over 3 T. Within T1MES repeatability for native T1 was narrow for several sequences, for example, at 1.5 T, Siemens MOLLI 5s(3s)3s prototype number 448B (mean CoV = 0.27%) and Philips modified Look-Locker inversion recovery (MOLLI) 3s(3s)5s (CoV 0.54%), and at 3 T, Philips MOLLI 3b(3s)5b (CoV 0.33%) and Siemens shortened MOLLI (ShMOLLI) prototype 780C (CoV 0.69%). After adjusting for temperature and field strength, it was found that the T1 mapping sequence and scanner software version (both P < 0.001 at 1.5 T and 3 T), and to a lesser extent the scanner model (P = 0.011, 1.5 T only), had the greatest influence on T1 across multiple centers. Conclusion: The T1MES CE/FDA approved phantom is a robust quality assurance device. In a multi-center setting, T1 mapping had performance differences between field strengths, sequences, scanner software versions, and manufacturers. However, several specific combinations of field strength, sequence, and scanner are highly repeatable, and thus, have potential to provide standardized assessment of T1 times for clinical use, although temperature correction is required for native T1 tubes at least. [ABSTRACT FROM AUTHOR]
Copyright of Journal of Cardiovascular Magnetic Resonance (BioMed Central) is the property of BioMed Central and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
قاعدة البيانات: Complementary Index
الوصف
تدمد:1532429X
DOI:10.1186/s12968-020-00613-3